Pagina 2

 EPA COMPLIANCEThe SMART3 Colorimeter is an EPA-Accepted instrument. EPA-Accepted means that the instrument meets the requirements for instrumentati

Pagina 3 - CONTENTS

PROCEDURE1. This test requires a Reagent Blank. Rinse a tube (0290) with clear, colorless, boron free water. Fill to 10 mL line with clear, colorless

Pagina 4

BROMINEDPD TABLET METHOD • CODE 3643-SCQUANTITY CONTENTS CODE100 *DPD #1 Instrument Grade Tablets *6903A-J100 *DPD #3 Instrument Grade Tablets *61

Pagina 5 - GENERAL INFORMATION

PROCEDURE A: BROMINE (NO CHLORINE)1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or anothe

Pagina 6 -  REGISTER YOUR METER

PROCEDURE C: FREE AVAILABLE, TOTAL AVAILABLE & COMBINED CHLORINE IN THE PRESENCE OF BROMINE1. Perform the test for free and combined chlorine as

Pagina 7 -  SPECIFICATIONS

BROMINE, DPD Tablet SMART3 Test Procedures 11.10Test Procedures

Pagina 8 - 6 SMART3 Colorimeter 11.10

BROMINE - UDVDPD METHOD–UNIT DOSE VIALS · CODE 4311-JQUANTITY CONTENTS CODE1 *Free Chlorine Unit Dose Vials, 20 pouches *4311-J*WARNING: Reagents

Pagina 9 - ACCESSORIES

APPLICATION: Drinking, surface, and saline waters; swimming pool water; domestic and industrial waters and wastes.RANGE: 0.0–20.0 ppm BromineMDL: 0.25

Pagina 10 -  IP67 CERTIFICATION

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another seq

Pagina 11 - CHEMICAL TESTING

BROMINE, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 12 -  FILTRATION

CADMIUMPAN METHOD • CODE 4017-01QUANTITY CONTENTS CODE60 mL *Buffered Ammonia Reagent *4020-H15 mL Sodium Citrate, 10% 6253-E30 mL *PAN Indica

Pagina 13 - SMART3 Colorimeter 11.10 11

CHEMICAL TESTING  WATER SAMPLING FOR CHEMICAL ANALYSISTaking Representative SamplesThe underlying factor to be considered for any type of water sampl

Pagina 14 -  METER CARE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 012

Pagina 15 -  CALIBRATION CURVES

CHLORIDEARGENTOMETRIC METHOD · CODE 3693-SC QUANTITY CONTENTS CODE50 *Chloride Spectrophotometric Grade Tablets *3885A-H1 Tablet Crusher 0175*

Pagina 16 - 14 SMART3 Colorimeter 11.10

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 020

Pagina 17 - SMART3 Colorimeter 11.10 15

CHLORINELIQUID DPD METHOD · CODE 4859QUANTITY CONTENTS CODE30 mL DPD 1A Free Chlorine Reagent P-6740-G30 mL *DPD 1B Free Chlorine Reagent *P-6

Pagina 18 -  STANDARD ADDITIONS

APPLICATION: Drinking, surface, and saline waters; swimming pool water; domestic and industrial wastes.RANGE: 0.00–4.00 ppm ChlorineMDL: 0.03 ppmMETHO

Pagina 19 - MEASUREMENTS

PROCEDURE–FREE CHLORINE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence

Pagina 20 -  STRAY LIGHT INTERFERENCE

CHLORINE, DPD Liquid SMART3 Test Procedures 11.10Test Procedures

Pagina 21 - SMART3 COLORIMETER

CHLORINEDPD TABLET METHOD • CODE 3643-SCQUANTITY CONTENTS CODE100 *DPD #1 Instrument Grade Tablets *6903A-J100 *DPD #3 Instrument Grade Tablets *6

Pagina 22 -  COMPONENTS

APPLICATION: Drinking, surface, and saline waters; swimming pool water; domestic and industrial wastes.RANGE: 0.00–4.00 ppm ChlorineMDL: 0.03 ppmMETHO

Pagina 23 - GENERAL OPERATING PROCEDURES

PROCEDURE–FREE CHLORINE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence

Pagina 24 - 22 SMART3 Colorimeter 11.10

Sampler (LaMotte Code 1060) will serve as a subsurface or in-depth sampler. This weighted device is lowered to the sampling depth and allowed to rest

Pagina 25 - Result or

CHLORINE, DPD Tablet SMART3 Test Procedures 11.10Test Procedures

Pagina 26 -  LOOPING MENUS

CHLORINE, FREE - UDVDPD METHOD–UNIT DOSE VIALS · CODE 4311-JQUANTITY CONTENTS CODE1 *Free Chlorine Unit Dose Vials, 20 pouches *4311-J*WARNING: Re

Pagina 27 -  TESTING MENU

Chlorine can be added in the form of chlorine gas, liquid sodium hypochlorite (bleach), granular calcium hypochlorite or as organic chlorine compounds

Pagina 28 -  GENERAL TESTING PROCEDURES

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another seq

Pagina 29

CHLORINE, Free, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 30 - 1.00 ppm

CHLORINE, TOTAL - UDVDPD METHOD–UNIT DOSE VIALS · CODE 4312-JQUANTITY CONTENTS CODE1 *Total Chlorine Unit Dose Vials, 20 pouches *4312-J*WARNING:

Pagina 31 - SMART3 Colorimeter 11.10 29

Chlorine can be added in the form of chlorine gas, liquid sodium hypochlorite (bleach), granular calcium hypochlorite or as organic chlorine compounds

Pagina 32 - 0.28 ppm

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another seq

Pagina 33 - 0.30 ppm

CHLORINE, Total, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 34 - 32 SMART3 Colorimeter 11.10

CHLORINE DIOXIDEDPD METHOD • CODE 3644-SCQUANTITY CONTENTS CODE100 *DPD #1 Instrument Grade Tablets *6903A-J15 mL Glycine Solution 6811-E1 T

Pagina 35

 AN INTRODUCTION TO COLORIMETRIC ANALYSISMost test substances in water are colorless and undetectable to the human eye. To test for their presence w

Pagina 36

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 019

Pagina 37 - SMART3 Colorimeter 11.10 35

CHROMIUM, HEXAVALENTDIPHENYLCARBOHYDRAZIDE METHODCODE 3645-SCQUANTITY CONTENTS CODE10 g *Chromium Reagent Powder *V-6276-D1 Spoon, 0.1 g, plasti

Pagina 38 - EDITING MENU

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 021

Pagina 39 - Editing/Set Up

CHROMIUM - HEXAVALENT, TRIVALENT & TOTALDIPHENYLCARBOHYDRAZIDE METHODCODE 3698-SCQUANTITY CONTENTS CODE60 mL *Sulfuric Acid, 5N *7681-H10 g *C

Pagina 40 - ADDING A TEST

APPLICATION: Drinking, surface, & saline water; domestic and industrial waste.RANGE: 0.00–1.00 ppm ChromiumMDL: 0.01 ppmMETHOD: The trivalent chro

Pagina 41

HEXAVALENT CHROMIUM PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 42 - DELETING A TEST

6. Remove fl ask from heat. Cool sample under running water. This is the digested sample.7. Pour digested sample into clean graduated cylinder (0418).

Pagina 43 - SMART3 Colorimeter 11.10 41

COBALTPAN METHOD · CODE 4851QUANTITY CONTENTS CODE60 mL *Cobalt Buffer *4852-H60 mL *Cobalt Indicator Reagent *4853-H30 mL *Stabilizer Solution *4

Pagina 44 -  EDIT USER TESTS

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 023

Pagina 45

COD – LOW RANGEMERCURY FREE DIGESTION METHOD • CODE 0072-SCMERCURY DIGESTION METHOD • CODE 0075-SCQUANTITY CONTENTS CODE25 *COD Low Range Mercury

Pagina 46 - H I J K

 REAGENT BLANKSome tests will provide greater accuracy if a reagent blank is determined to compensate for any color or turbidity resulting from the

Pagina 47 - O P Q R

APPLICATION: Domestic and industrial wastes.RANGE: 0–150 mg/L CODMDL: 7.5 mg/LMETHOD: Dichromate in the presence of silver salts, at high temperature

Pagina 48 - 46 SMART3 Colorimeter 11.10

PROCEDUREUse COD/UDV adapter.1. Homogenize sample if necessary.2. Preheat COD heater block to 150±2°C.3. Remove cap from COD tube. Hold tube at a 4

Pagina 49

acid.For greater accuracy, a minimum of three repetitions should be performed and the results averaged.Some samples may be digested completely in less

Pagina 50 - 1 1 2 3

COD – STANDARD RANGEMERCURY FREE DIGESTION METHOD • CODE 0073-SCMERCURY DIGESTION METHOD • CODE 0076-SCQUANTITY CONTENTS CODE25 *COD Standard Rang

Pagina 51 - . – 0 1

APPLICATION: Domestic and industrial wastes.RANGE: 0–1500 mg/L CODMDL: 40 mg/LMETHOD: Dichromate in the presence of silver salts, at high temperature

Pagina 52

PROCEDUREUse COD/UDV adapter.1. Homogenize sample if necessary.2. Preheat COD heater block to 150±2°C.3. Remove cap from COD tube. Hold tube at a 4

Pagina 53 - SMART3 Colorimeter 11.10 51

acid.For greater accuracy, a minimum of three repetitions should be performed and the results averaged.Some samples may be digested completely in less

Pagina 54 - 52 SMART3 Colorimeter 11.10

COD – HIGH RANGEMERCURY FREE DIGESTION METHOD • CODE 0074-SCMERCURY DIGESTION METHOD • CODE 0077-SCQUANTITY CONTENTS CODE25 *COD High Range Mercur

Pagina 55 - . – 0 1

APPLICATION: Domestic and industrial wastes.RANGE: 0–15000 mg/L CODMDL: 400 mg/LMETHOD: Dichromate in the presence of silver salts, at high temperatur

Pagina 56

PROCEDUREUse COD/UDV adapter.1. Homogenize sample if necessary.2. Preheat COD heater block to 150±2°C.3. Remove cap from COD tube. Hold tube at a 4

Pagina 57

Below is a list of suggested wavelengths for the color of the reacted samples. Use these as a starting point.Sample Color Wavelength RangeYellow 428Pi

Pagina 58

acid.For greater accuracy, a minimum of three repetitions should be performed and the results averaged.COD, High Range SMART3 Test Procedures 11.10T

Pagina 59 -  SETTING THE CLOCK

COLORPLATINUM COBALT METHODNO REAGENTS REQUIREDColor in water may be attributed to humus, peat, plankton, vegetation, and natural metallic ions, such

Pagina 60 -  LOGGING DATA

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 027

Pagina 61 -  SETTING POWER SAVE

COPPER – LOW RANGEBICINCHONINIC ACID METHOD • CODE 3640-SCQUANTITY CONTENTS CODE50 *Copper Tablets *T-3808-H*WARNING: Reagents marked with an * ar

Pagina 62 - 60 SMART3 Colorimeter 11.10

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 028

Pagina 63 -  SETTING THE BACKLIGHT TIME

COPPERCUPRIZONE METHOD • CODE 4023QUANTITY CONTENTS CODE15 mL Copper A P-6367-E15 mL *Copper B *P-6368-E*WARNING: Reagents marked with an * ar

Pagina 64 -  SELECTING A LANGUAGE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 030

Pagina 65 - SMART3 Colorimeter 11.10 63

COPPERDIETHYLDITHIOCARBAMATE METHOD • CODE 3646-SCQUANTITY CONTENTS CODE15 mL *Copper 1 *6446-E*WARNING: Reagents marked with an * are considered

Pagina 66 - COMPUTER CONNECTION

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 031

Pagina 67 -  BATTERY/AC OPERATION

COPPER – UDVBICINCHONINIC ACID METHOD–UNIT DOSE VIALSCODE 4314-JQUANTITY CONTENTS CODE1 Copper Unit Dose Vials, 20 pouches 4314-JEquipment needed but

Pagina 68 - MAINTENANCE

to scan blank.3. Following the individual reagent system instructions, react each standard solution beginning with 0 ppm. Continue with standards in

Pagina 69 - TROUBLESHOOTING

APPLICATION: Drinking, surface, and saline waters; domestic and industrial wastes.RANGE: 0.0–4.0 ppm CopperMDL: 0.1 ppmMETHOD: Cupric ions form a purp

Pagina 70 -  TROUBLESHOOTING GUIDE

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 71 - 02.13.12

COPPER, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 72

CYANIDEPYRIDINE-BARBITURIC ACID METHODCODE 3660-01-SCQUANTITY CONTENTS CODE60 mL Cyanide Buffer 2850PS-H5 g *Cyanide Cl Reagent *2794DS-C5 g

Pagina 73 - REAGENT SYSTEMS

APPLICATION: Low level concentrations in drinking and surface waters; domestic and industrial waters. This method determines only those cyanides amena

Pagina 74

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to selct TESTING MENU.3. Select ALL TESTS (or another sequence containing 034

Pagina 75

CYANIDE SMART3 Test Procedures 2.11Test Procedures

Pagina 76

CYANURIC ACIDMELAMINE METHOD–TURBIDITY • CODE 3661-01-SCQUANTITY CONTENTS CODE2 x 100 mL *Cyanuric Acid Test Solution *4856-J1 Syringe, 5 mL 0

Pagina 77 - ALKALINITY–UDV

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 035

Pagina 78

CYANURIC ACID – UDVMELAMINE METHOD-TURBIDITY–UNIT DOSE VIALSCODE 4313-JQUANTITY CONTENTS CODE1 Cyanuric Acid Unit Dose Vials, 20 pouches 4313-JEquipme

Pagina 79 - ALUMINUM

PREPARING DILUTE STANDARD SOLUTIONSStandard solutions should be prepared to create a calibration curve. Standard solutions can be prepared by diluting

Pagina 80

APPLICATION: Swimming pool water.RANGE: 5–150 ppm Cyanuric AcidMDL: 10 ppmMETHOD: A buffered solution of melamine forms a precipitate with cyanuric ac

Pagina 81 - AMMONIA NITROGEN - LOW RANGE

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to selct TESTING MENU.3. Select ALL TESTS (or another sequ

Pagina 82 - PROCEDURE - SALT WATER

CYANURIC ACID, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 83

DISSOLVED OXYGENWINKLER COLORIMETRIC METHOD • CODE 3688-SCQUANTITY CONTENTS CODE30 mL *Manganese Sulfate Solution *4167-G30 mL *Alkaline Potassium

Pagina 84 - AMMONIA NITROGEN - HIGH RANGE

COLLECTION & TREATMENT OF THE WATER SAMPLESteps 1 through 4 below describe proper sampling technique in shallow water. For sample collection at de

Pagina 85

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 038

Pagina 86

DISSOLVED OXYGEN SMART3 Test Procedures 11.10Test Procedures

Pagina 87

FLUORIDESPADNS METHOD • CODE 3647-02-SCQUANTITY CONTENTS CODE4 x 30 mL *Acid Zirconyl SPADNS Reagent *3875-G2 x 30 mL *Sodium Arsenite Solution *4

Pagina 88

The following substances produce a negative interference at the concentration given:Alkalinity (CaCO3) 5000 mg/LAluminum (Al3+) 0.1 mg/LIron (Fe3+)

Pagina 89 - BENZOTRIAZOLE/TOLYLTRIAZOLE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 040

Pagina 90

 STANDARD ADDITIONSA common method to check the accuracy and precision of a test is by standard additions. In this method a sample is tested to dete

Pagina 91 - BENZOTRIAZOLE PROCEDURE

FLUORIDE SMART3 Test Procedures 11.10Test Procedures

Pagina 92 - TOLYLTRIAZOLE PROCEDURE

HARDNESS, TOTAL – UDVUNIT DOSE VIALS • CODE 4309-JQUANTITY CONTENTS CODE1 Calcium Hardness Unit Dose Vials, 20 pouches 4309-JEquipment needed but not

Pagina 93 - BIGUANIDE

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 94

HYDRAZINEp-DIMETHYLAMINOBENZALDEHYDE METHOD CODE 3656-01-SCQUANTITY CONTENTS CODE2 x 60 mL *Hydrazine Reagent A *4841-H10 g *Hydrazine Reagent B P

Pagina 95 - BORATE - UDV

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 045

Pagina 96

HYDROGEN PEROXIDE – LOW RANGEDPD METHOD • CODE 3662-SCQUANTITY CONTENTS CODE30 mL *Hydrogen Peroxide Reagent #1 *6452-G2 x 100 *Hydrogen Peroxide

Pagina 97

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence containing 046

Pagina 98

HYDROGEN PEROXIDE – HIGH RANGEDPD METHOD • CODE 4045-01QUANTITY CONTENTS CODE30 mL *Hydrogen Peroxide Reagent #1 *6452-G2 x 100 *Hydrogen Peroxide

Pagina 99

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence containing 047

Pagina 100 - PROCEDURE

HYDROGEN PEROXIDE – SHOCKDPD METHOD • CODE 4045-01QUANTITY CONTENTS CODE30 mL *Hydrogen Peroxide Reagent #1 *6452-G2 x 100 *Hydrogen Peroxide LR T

Pagina 101 - Test Procedures

 SAMPLE DILUTION TECHNIQUES & VOLUMETRIC MEASUREMENTSIf a test result using the SMART3 Colorimeter gives an over range message then the the samp

Pagina 102 - PROCEDURE B:

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence containing 048

Pagina 103 - PROCEDURE C:

IODINEDPD METHOD TABLET • CODE 3643-SCQUANTITY CONTENTS CODE100 *DPD #1 Instrument Grade Tablets *6903A-J100 *DPD #3 Instrument Grade Tablets *619

Pagina 104

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 050

Pagina 105 - BROMINE - UDV

IRONBIPYRIDYL METHOD • CODE 3648-SCQUANTITY CONTENTS CODE30 mL *Iron Reagent #1 *V-4450-G5 g *Iron Reagent #2 Powder *V-4451-C1 Pipet, 0.5 mL, p

Pagina 106

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 051

Pagina 107

IRON – UDVBIPYRIDYL METHOD–UNIT DOSE VIALS · CODE 4315-JQUANTITY CONTENTS CODE1 *Total Iron Unit Dose Vials, 20 pouches *4315-J*WARNING: Reagents

Pagina 108

APPLICATION: Drinking, surface, and saline waters; domestic and industrial wastes.RANGE: 0.00–10.00 ppmMDL: 0.05 ppmMETHOD: Ferric iron is reduced to

Pagina 109 - PAN METHOD • CODE 4017-01

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 110

IRON, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 111 - CHLORIDE

IRONI,I0-PHENANTHROLINE METHOD • CODE 3668-SCQUANTITY CONTENTS CODE15 mL *Acid Phenanthroline Indicator *2776-E5 g *Iron Reducing Reagent *2777-C1

Pagina 113 - CHLORINE

 INTERFERENCESLaMotte reagent systems are designed to minimize most common interferences. Each individual test instruction discusses interferences u

Pagina 114

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 053

Pagina 115 - PROCEDURE–TOTAL CHLORINE

LEADPAR METHOD · CODE 4031QUANTITY CONTENTS CODE250 mL *Ammonium Chloride Buffer *4032-K15 mL *Sodium Cyanide, 10% *6565-E30 mL *PAR Indicator *40

Pagina 116

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 054

Pagina 117

MANGANESE – LOW RANGEPAN METHOD • CODE 3658-01-SCQUANTITY CONTENTS CODE2x60 mL *Hardness Buffer Reagent *4255-H30 mL *Manganese Indicator Reagent

Pagina 118

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 055

Pagina 119 - PROCEDURE–COMBINED CHLORINE

MANGANESE – HIGH RANGEPERIODATE METHOD • CODE 3669-SCQUANTITY CONTENTS CODE10 g Manganese Buffer Reagent 6310-D15 g *Manganese Periodate Reage

Pagina 120

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 056

Pagina 121 - CHLORINE, FREE - UDV

MERCURYTMK METHOD · CODE 4861QUANTITY CONTENTS CODE50 *TMK Tablets *4862-H2 x 250 mL *Propyl Alcohol *4863-K250 mL *Acetate Buffer *4864-K1 Tabl

Pagina 122

PREPARATION OF *TMK INDICATORNOTE: Prepare *TMK Indicator daily. Keep out of direct sunlight.1. Fill test tube (0778) to the 10 mL line with *Propyl

Pagina 123

MOLYBDENUM – HIGH RANGETHIOGLYCOLATE METHOD • CODE 3699-03-SCQUANTITY CONTENTS CODE2 x 30 mL *Mo Buffer *3997-G2 x 30 mL *Molybdenum Oxidizing Reag

Pagina 124

OPERATION OF THE SMART3 COLORIMETER  OVERVIEWThe SMART3 is a portable, microprocessor controlled, direct reading colorimeter. It has a graphical liqu

Pagina 125 - CHLORINE, TOTAL - UDV

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 061

Pagina 126

NICKELDIMETHYLGLYOXIME METHOD • CODE 3663-SCQUANTITY CONTENTS CODE60 mL *Hydrochloric Acid, 2.5N *6251PS-H30 g *Ammonium Persulfate Reagent *6566-

Pagina 127

PROCEDURE1. Use the 10 mL graduated cylinder (0416) to measure 10 mL of sample water. Pour into glass test tube (0608).2. Use the 1 mL pipet (0354)

Pagina 128

NITRATEZINC REDUCTION · CODE 3689-SCQUANTITY CONTENTS CODE50 *Nitrate Spectrophotometric Grade Tablets *3881A-H1 Tablet Crusher 0175*WARNING:

Pagina 129 - CHLORINE DIOXIDE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 065

Pagina 130

NITRATE - UDVZINC REDUCTION METHOD-UNIT DOSE VIALS •CODE 4321-J QUANTITY CONTENTS CODE50 Nitrate UDV 4321-JEquipment needed but not supplied:STANDARD

Pagina 131 - CHROMIUM, HEXAVALENT

APPLICATION: swimming pools, drinking water, surface and saline watersRANGE: 0.00 – 80.0 ppm nitrateMDL: 2 ppmMETHOD: Zinc is used to reduce nitrate t

Pagina 132

PROCEDUREUse COD/UDV adapter.1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 133 - TRIVALENT & TOTAL

NITRATE, UDV SMART3 Test Procedures 1.11Test Procedures

Pagina 134

NITRATE-NITROGEN – LOW RANGECADMIUM REDUCTION METHOD • CODE 3649-SCQUANTITY CONTENTS CODE2 x 60 mL *Mixed Acid Reagent *V-6278-H5 g *Nitrate Reduc

Pagina 135 - HEXAVALENT CHROMIUM PROCEDURE

 COMPONENTSFigure 1 shows a diagram of the SMART3 Colorimeter and its components.20 SMART3 Colorimeter 11.10

Pagina 136 - TRIVALENT CHROMIUM PROCEDURE

APPLICATION: This method determines nitrate levels in drinking, surface, saline waters, domestic and industrial waters.RANGE: 0.00–3.00 ppm Nitrate Ni

Pagina 137 - PAN METHOD · CODE 4851

PROCEDURENOTE: Place Dispenser Cap (0692) on *Mixed Acid Reagent (V-6278). Save this cap for refi ll reagents.1. Press and hold until colorimeter tu

Pagina 138

NITRATE NITROGEN, Low Range SMART3 Test Procedures 11.10Test Procedures

Pagina 139 - COD – LOW RANGE

NITRITE-NITROGEN – LOW RANGEDIAZOTIZATION METHOD • CODE 3650-SCQUANTITY CONTENTS CODE2 x 60 mL *Mixed Acid Reagent *V-6278-H5 g *Color Developing

Pagina 140

PROCEDURENOTE: Place Dispenser Cap (0692) on *Mixed Acid Reagent (V-6278). Save this cap for refi ll reagents.1. Press and hold until colorimeter tu

Pagina 141

NITROGEN, TOTALCHROMOTROPIC ACID WITH PERSULFATE DIGESTION METHOD • CODE 4026-01QUANTITY CONTENTS CODE25 Total Nitrogen Hydroxide Reagent Tubes

Pagina 142

in conjunction with phosphate stimulate the growth of algae with all of the related diffi culties associated with excessive algae growth.U.S. Public He

Pagina 143 - COD – STANDARD RANGE

PROCEDUREUse COD/UDV adapter.1. Preheat COD reactor to 100 ±2°C. Follow safety precautions.2. Remove caps from two *Total Nitrogen Hydroxide Reagent

Pagina 144

other Total Nitrogen Acid Reagent Tube. This is the sample.22. Cap the tubes and invert 10 times to mix. CAUTION: The tubes will be hot.Note: Invert

Pagina 145

OXYGEN SCAVENGERSDEHA (DIETHYLHYDROXYLAMINE), CARBOHYDRAZIDE, ERYTHORBIC ACID, HYDROQUINONE, METHYLETHYLKETOXIMEIRON REDUCTION METHOD · CODE 4857QUANT

Pagina 146

GENERAL OPERATING PROCEDURESThe operation of the SMART3 Colorimeter is controlled by a microprocessor. The microprocessor is programmed with menu driv

Pagina 147 - COD – HIGH RANGE

APPLICATION: BoilersRANGE: 0.000–0.700 ppm DEHA (Diethylhydroxylamine)0.000–0.900 ppm Carbohydrazide0.00–3.00 ppm Erythorbic Acid0.00–2.00 ppm Hydroqu

Pagina 148

DEHA PROCEDURE1. Press and hold until colorimeter turns on.2. PressENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing

Pagina 149

CARBOHYDRAZIDE PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence

Pagina 150

ERYTHORBIC ACID PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequenc

Pagina 151 - NO REAGENTS REQUIRED

HYDROQUINONE PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence c

Pagina 152

METHYLETHYLKETOXIME PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 153 - COPPER – LOW RANGE

OXYGEN SCAVENGERS SMART3 Test Procedures 11.10Test Procedures

Pagina 154

OZONEDPD METHOD • CODE 4881QUANTITY CONTENTS CODE30 mL DPD #1A Free Chlorine Reagent P-6740-G30 mL *DPD #1B Free Chlorine Reagent *P-6741-G30

Pagina 155 - CUPRIZONE METHOD • CODE 4023

PROCEDURE - OZONE ONLY1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence c

Pagina 156

PROCEDURE - OZONE WITH CHLORINE PRESENT1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or a

Pagina 157

 THE DISPLAY & THE MENUS The display allows menu selections to be viewed and selected. These selections instruct the SMART3 to perform specifi c

Pagina 158

OZONE, DPD, Liquid SMART3 Test Procedures 11.10Test Procedures

Pagina 159 - COPPER – UDV

OZONEINDIGO METHOD • CODE 3651-SCQUANTITY CONTENTS CODE15 mL Chlorine Inhibitor 3990-E250 mL *Ozone Buffer *3991-K30 mL Indigo Blue Stock So

Pagina 160

PROCEDURE–LOW RANGEA. PREPARATION OF HR REAGENTNOTE: The quantity of Indigo Blue Stock solution (3989) supplied will prepare one batch of HR Reagent f

Pagina 161

PROCEDURE–HIGH RANGEA. PREPARATION OF HR REAGENTNOTE: The quantity of Indigo Blue Stock solution (3989) supplied will prepare one batch of HR Reagent

Pagina 162

Test ProceduresOZONE, Indigo SMART3 Test Procedures 11.10

Pagina 163 - CODE 3660-01-SC

pHCOLORIMETRIC METHOD • CODE 3700-01-SCQUANTITY CONTENTS CODE60 mL Chlorphenol Red Indicator V-2209-H60 mL Phenol Red Indicator V-2304-H60 mL Thymol B

Pagina 164

APPLICATION: Drinking, surface, and saline waters, swimming pool water; domestic and industrial wastes.METHOD: The various pH indicators exhibit a spe

Pagina 165

PROCEDURE1. Use Indicator, Range, & Test Name chart to select the indicator, corresponding to anticipated pH range and to determine corresponding

Pagina 166

Test ProcedurespH SMART3 Test Procedures 11.10

Pagina 167 - CYANURIC ACID

PHENOLAMINOANTIPYRINE METHOD • CODE 3652-01-SCQUANTITY CONTENTS CODE5 g Aminoantipyrine Reagent 7825-C30 mL *Ammonium Hydroxide Solution *7826

Pagina 168

Header Menu TitleMessage or Result WindowResult or MessageAnotherMain Window Selection And AnotherFooter12:00:00 001/500

Pagina 169 - CYANURIC ACID – UDV

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 077

Pagina 170

PHOSPHATE – LOW RANGEASCORBIC ACID REDUCTION METHODCODE 3653-SCQUANTITY CONTENTS CODE60 mL *Phosphate Acid Reagent *V-6282-H5 g *Phosphate Reducin

Pagina 171

APPLICATION: Drinking, surface and saline waters; domestic and industrial wastes (Method based on reactions that are specifi c for orthophosphate).RANG

Pagina 172

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 078

Pagina 173 - DISSOLVED OXYGEN

PHOSPHATE, Low Range SMART3 Test Procedures 11.10Test Procedures

Pagina 174

PHOSPHATE – HIGH RANGEVANADOMOLYBDOPHOSPHORIC ACID METHODCODE 3655-SCQUANTITY CONTENTS CODE4 x 30 mL *VM Phosphate Reagent *4410-G1 Pipet, 1.0 m

Pagina 175

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 079

Pagina 176

PHOSPHATE ppbASCORBIC ACID REDUCTION METHODCODE 3653-SCQUANTITY CONTENTS CODE60 mL *Phosphate Acid Reagent *V-6282-H5 g *Phosphate Reducing Reagen

Pagina 177 - FLUORIDE

APPLICATION: Drinking, surface and saline waters; domestic and industrial wastes (Method based on reactions that are specifi c for orthophosphate).RANG

Pagina 178

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 080

Pagina 179

 LOOPING MENUSLong menus, such as All Tests, incorporate a looping feature which allows the user to quickly reach the last choice in the menu from t

Pagina 180

Test ProceduresPHOSPHATE, ppb SMART3 Test Procedures 11.10

Pagina 181 - HARDNESS, TOTAL – UDV

PHOSPHORUS, TOTAL - LOW RANGEASCORBIC ACID REDUCTION WITH PERSULFATE DIGESTION METHOD • CODE 4024-01QUANTITY CONTENTS CODE25 *Total Phosphorus Aci

Pagina 182

Phosphorus in natural waters and wastewaters occurs almost exclusively in the form of orthophosphates, condensed phosphates (pyro-, meta- and other po

Pagina 183 - HYDRAZINE

PROCEDUREUse COD/UDV adapter.1. Preheat COD reactor to 150 ±2°C. Follow safety precautions.2. Remove cap from a *Total Phosphorus Acid Reagent Tube

Pagina 184

PHOSPHORUS, Total, Low Range SMART3 Test Procedures 11.10Test Procedures

Pagina 185 - HYDROGEN PEROXIDE – LOW RANGE

PHOSPHORUS, TOTAL – HIGH RANGEMOLYBDOVANADATE METHOD WITH ACID PERSULFATE DIGESTION • CODE 4025-01QUANTITY CONTENTS CODE25 *Total Phosphorus Acid R

Pagina 186

Phosphorus in natural waters and wastewaters occurs almost exclusively in the form of orthophosphates, condensed phosphates (pyro-, meta- and other po

Pagina 187 - DPD METHOD • CODE 4045-01

PROCEDUREUse COD/UDV adapter.1. Preheat COD reactor to 150 ±2°C. Follow safety precautions.2. Remove cap from a *Total Phosphorus Acid Reagent Tube

Pagina 188

PHOSPHORUS, Total, High Range SMART3 Test Procedures 11.10Test Procedures

Pagina 189 - HYDROGEN PEROXIDE – SHOCK

POTASSIUMTETRAPHENYLBORON METHOD • CODE 3639-SCQUANTITY CONTENTS CODE30 mL *Sodium Hydroxide, 1.0N *4004WT-G5 g *Tetraphenylboron Powder *6364-

Pagina 190

TESTING  TESTING MENUThe Testing Menu is used to run all LaMotte pre-programmed tests, User Tests and Absorbance tests at one of four wavelengths. Te

Pagina 191

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 083

Pagina 192

SILICA – LOW RANGEHETEROPOLY BLUE METHOD • CODE 3664-SCQUANTITY CONTENTS CODE30 mL *Silica Reagent #1 *V-4466-G30 mL *Silica Reagent #2 *V-4467-G3

Pagina 193

APPLICATION: Drinking, surface and saline waters; domestic and industrial wastes.RANGE: 0.0–4.0 ppm SilicaMDL: 0.05 ppmMETHOD: Reactive silica forms a

Pagina 194

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 085

Pagina 195 - IRON – UDV

SILICA, Low Range SMART3 Test Procedures 11.10Test Procedures

Pagina 196

SILICA – HIGH RANGESILICOMOLYBDATE METHOD • CODE 3687-SCQUANTITY CONTENTS CODE30 mL *Silica Reagent #1 *V-4466-G30 mL *Silica Reagent #2 *V-4467-G

Pagina 197

1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 086 Silica H

Pagina 198

SULFATE – HIGH RANGEBARIUM CHLORIDE METHOD • CODE 3665-SCQUANTITY CONTENTS CODE10 g *Sulfate Reagent *V-6277-D1 Spoon, 0.1 g, plastic 0699*WAR

Pagina 199

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 089

Pagina 200

SULFIDE – LOW RANGEMETHYLENE BLUE METHOD • CODE 3654-02-SCQUANTITY CONTENTS CODE2 x 30 *Sulfi de Reagent A *V-4458-G15 mL *Sulfi de Reagent B *V-44

Pagina 201 - PAR METHOD · CODE 4031

 TEST SEQUENCESSequence 1, Sequence 2, And Sequence 3 are alterable sequences. They may be edited using the Editing Menu. Any of the LaMotte pre-pro

Pagina 202

APPLICATION: Drinking, surface and saline waters; domestic and industrial wastes.RANGE: 0.00–1.50 ppm Sulfi deMDL: 0.06 ppmMETHOD: Under suitable condi

Pagina 203 - MANGANESE – LOW RANGE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 090

Pagina 204

SULFIDE, Low Range SMART3 Test Procedures 11.10Test Procedures

Pagina 205 - MANGANESE – HIGH RANGE

SURFACTANTSION PAIR EXTRACTION–BROMPHENOL BLUE INDICATOR METHOD • CODE 4876-01QUANTITY CONTENTS CODE50 g pH Adjustment Powder 4509- H10 g So

Pagina 206

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 094

Pagina 207 - TMK METHOD · CODE 4861

TANNINTUNGSTO-MOLYBDOPHOSPHORIC ACID METHOD CODE 3666-01-SC QUANTITY CONTENTS CODE30 mL *Tannin Reagent #1 *7833-G2 x 60 mL *Tannin Reagent #2 *78

Pagina 208 - PREPARATION OF *TMK INDICATOR

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 096

Pagina 209 - MOLYBDENUM – HIGH RANGE

TURBIDITYABSORPTION METHOD • NO REAGENTS REQUIREDTurbidity is a measure of water clarity and is independent of color. Turbidity is caused by undissolv

Pagina 210

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 098

Pagina 211

ZINC – LOW RANGEZINCON METHOD • CODE 3667-SCQUANTITY CONTENTS CODE30 mL *Zinc Indicator Solution *6314-G120 mL *Methyl Alcohol *6319-J10 g Sodiu

Pagina 212

 TESTING WITH LaMOTTE PRE-PROGRAMMED TESTS1. Press and briefl y hold to turn the meter on. The LaMotte logo screen will appear for about 3 seconds

Pagina 213

APPLICATION: Drinking and surface waters, domestic and industrial waste water.RANGE: 0.00–3.00 ppm ZincMDL: 0.05 ppmMETHOD: Zinc forms a blue colored

Pagina 214

PROCEDUREA. PREPARATION OF DILUTE ZINC INDICATOR SOLUTION1. Use a pipet (0352) to add exactly 5.0 mL of *Zinc Indicator Solution (6314) to a 10 mL gr

Pagina 215 - NITRATE - UDV

APPENDIXAmmonia in water occurs in two forms: toxic unionized ammonia (NH3) and the relatively non-toxic ionized form, ammonium ion (NH4+). This test

Pagina 216

FOR EXAMPLE:If a fresh water sample at 20°C has a pH of 8.5 and the test result is 1.0 ppm as Total Ammonia-Nitrogen:1. The percentage from the table

Pagina 217

AppendixAPPENDIX SMART3 Test Procedures 11.10

Pagina 218

CONTENTS  GENERAL INFORMATIONPackaging & Delivery ... 3General

Pagina 219 - NITRATE-NITROGEN – LOW RANGE

6. Insert the blank into the chamber. Close the lid.Press ENTER to scan the blank. The screen wil display Blank Done for about 1 second and then retu

Pagina 220

 CALIBRATING LaMOTTE PRE-PROGRAMMED TESTSThe LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of the pre-programmed tests by the

Pagina 221

4. Press or to scroll to the desired test factor.All Tests001 Alkalinity UDV002 Aluminum003 Ammonia-N LRF004 Ammonia-N LRS12:00:00

Pagina 222

8. The displayed result can now be standardized. Press or to scroll to calibrate.002 Aluminum0.28 ppm%T/AbsCalibrate12:00:00 001

Pagina 223 - NITRITE-NITROGEN – LOW RANGE

12. Press ENTER to select Set Calibration and save the calibration. Or press to scroll to Factory Setting. Press ENTER to select Factory Setting to

Pagina 224

 MEASURING IN THE ABSORBANCE MODE1. Press and briefl y hold to turn the meter on. The LaMotte logo screen will appear for about 3 seconds and the

Pagina 225 - NITROGEN, TOTAL

5. Press or to scroll to desired wavelength.Absorbance101 Absorbance 428102 Absorbance 525103 Absorbance 568104 Absorbance 63512:00:00

Pagina 226

9. To repeat the test, press ENTER to scan the sample again. The last blank scanned is used by the colorimeter for repeated scans. A different blank

Pagina 227

EDITING MENUThe Editing Menu allows the user to edit sequences, edit user tests, set the clock, edit the logging function, access factory setting, set

Pagina 228

4. Press ENTER to select Edit Sequences.Edit SequencesEdit Sequence 1Edit Sequence 2Edit Sequence 312:00:00 001/500 5

Pagina 229 - OXYGEN SCAVENGERS

Test Sequences ... 26General Testing Procedures ...

Pagina 230

 ADDING OR DELETING A TESTThere are three ways to alter a sequence: Insert Before, Insert After, and Delete. Insert Before adds a new test to the se

Pagina 231 - DEHA PROCEDURE

5. Press or to scroll to the test that will be added to the sequence. In this example, Aluminum.All Tests001 Alkalinity UDV002 Aluminum003 Ammonia

Pagina 232 - CARBOHYDRAZIDE PROCEDURE

DELETING A TESTBelow is a step-by-step example of how to delete a test in SEQUENCE 2 starting from the EDIT SEQUENCE 2 menu.1. To delete a test, pre

Pagina 233 - ERYTHORBIC ACID PROCEDURE

5. Press EXIT to exit the Edit Sequence menu and return to the Editing Menu.Editing MenuEdit SequencesEdit User TestSet ClockLogging12:00:00

Pagina 234 - HYDROQUINONE PROCEDURE

 EDIT USER TESTSIf a test other than the LaMotte programmed tests is performed regularly, a calibration for it may be entered in one of the 25 User

Pagina 235 - METHYLETHYLKETOXIME PROCEDURE

3. Press ENTER to select Editing Menu. Press to scroll to Edit User Test.Editing MenuEdit SequencesEdit User TestSet ClockLogging12:00:00

Pagina 236

 NAMING THE TESTA User Test can be up to 16 characters long. The menu choices for each character are 26 upper case letters A to Z, 26 lower case let

Pagina 237 - DPD METHOD • CODE 4881

4. Press ENTER to save the character and move to the next character.Name the Test P Q R S T U V111 H S ER TEST 0712:00:00

Pagina 238 - PROCEDURE - OZONE ONLY

8. Press ENTER to save the character. Repeat the procedure until the test name is complete. To remove a character, change the character to a space (lo

Pagina 239

 SELECTING THE VIAL AND WAVELENGTHThe SMART3 Colorimeter accepts three different vials (the 25 mm 0290 tube, UDVs and COD tubes) at 4 different wave

Pagina 240

GENERAL INFORMATION  PACKAGING & DELIVERYExperienced packaging personnel at LaMotte Company assure adequate protection against normal hazards enc

Pagina 241 - INDIGO METHOD • CODE 3651-SC

 ENTERING A TWO POINT CALIBRATIONThe SMART3 Colorimeter can scan two reacted standards and create a calibration curve. To prepare a calibration curv

Pagina 242 - PROCEDURE–LOW RANGE

4. Press ENTER to save the character and move to the next character.Low Standard 7 8 9 . – 0 11 . 00000012:00:00 001/5

Pagina 243 - PROCEDURE–HIGH RANGE

8. Press ENTER to save the character. Repeat the procedure until the low concentration value is complete. After the fi nal character is complete the me

Pagina 244

11. Insert the blank. Press ENTER. The screen will display Blank Done... for about 1 second and the Insert Low Standard screen will appear.Insert Low

Pagina 245

 ENTERING A MULTIPLE POINT CALIBRATIONThe SMART3 can directly create a 2 point calibration curve. (See Entering a Two Point Calibration on page 47.)

Pagina 246

3. Press ENTER to begin entering the values for the constants. A reverse font (dark background with a light character) will appear to indicate the cha

Pagina 247

8. Press ENTER to save the character. Repeat the procedure until the K0 value is complete. After the fi nal character is complete the meter will save t

Pagina 248

 SELECTING THE NUMERICAL FORMAT OF THE RESULTTo input tests with very different ranges, the number of decimal places displayed for a result can be s

Pagina 249

 SELECTING THE UNITS OF CONCENTRATIONThe SMART3 Colorimeter has seven options for units of concentration. They are No Units, ppm, FAU, pH, ppb, ppt

Pagina 250

 SETTING THE CLOCKSetting the clock allows the correct time and date stamp to be stored with each reading in the data logger.1. From the Editing Men

Pagina 251 - PHOSPHATE – LOW RANGE

 LIMITS OF LIABILITYUnder no circumstances shall LaMotte Company be liable for loss of life, property, profi ts, or other damages incurred through th

Pagina 252

 LOGGING DATAThe default setting for the data logger is enabled. The meter will log the last 500 data points. The counter in the center bottom of th

Pagina 253

5. Press EXIT to return to the Editing Menu.Editing MenuEdit SequenceEdit User TestSet ClockLogging12:00:00 001/500

Pagina 254

4. Press ENTER. The screen will display Storing... for about 1 second and the meter will return to the Editing Menu.Editing MenuSet ClockLoggingFactor

Pagina 255 - PHOSPHATE – HIGH RANGE

 SETTING THE BACKLIGHT TIMEThe backlight illuminates the display for enhanced viewing. The default setting is 10 seconds. If Button Control is chose

Pagina 256

 SELECTING A LANGUAGEThere are seven languages available in the SMART3: English, Spanish, French, Portuguese, Italian, Chinese, and Japanese.1. From

Pagina 257 - PHOSPHATE ppb

NOTE: If meter unintentionally switches to another language, use the procedure above to reset the meter to the desired language. For example, to reset

Pagina 258

COMPUTER CONNECTION  PC LINKThe SMART3 may be interfaced with any Windows-based computer by using the LaMotte SMARTLink 3 Program and USB Cable. The

Pagina 259

BATTERY  BATTERY/AC OPERATIONThe SMART3 may be operated on battery power, using a USB wall adapter or USB computer connection. If using the meter as

Pagina 260

less capacity after hundreds of charge cycles. This unit uses a custom battery assembly that is only available from LaMotte Company. Battery replacem

Pagina 261 - PHOSPHORUS, TOTAL - LOW RANGE

TROUBLESHOOTING  ERROR MESSAGESOVER RANGEIf the message OVERRANGE is displayed when scanning a sample, the sample may be over range or under range.

Pagina 262

 SPECIFICATIONSINSTRUMENT TYPE: ColorimeterReadout 160 x 100 backlit LCD, 20 x 6 line graphical displayWavelengths 428 nm, 525 nm, 568 nm, 635 nmWav

Pagina 263

 TROUBLESHOOTING GUIDEPROBLEM REASON SOLUTION FlashingLow battery. Readings are reliable.Charge battery or use USB wall/computer adapter.“Low Batter

Pagina 264

1910-TEST02.13.12

Pagina 266

SMART3 COLORIMETER REAGENT SYSTEMSSMART3 REAGENT SYSTEMS LISTLaMotte Company continuously updates the list of pre-programmed tests as the calibrations

Pagina 267

Cobalt (023) 0.00–2.00 0.04 PAN (3) 50COD-Low Range (024) 0–150 7.5 Digestion (1) 25COD-Standard Range (025) 0-1500 40 Digestion (1) 25COD-High Range

Pagina 268

Nitrate Nitrogen-Low Range (064) 0.00–3.00 0.10 Cadmium Reduction (2) 20Nitrate TesTab (065) 0-60 5 Zinc Reduction (1) 50Nitrate-UDV (066) 0.00-80.0 2

Pagina 270

ALKALINITY–UDVUNIT DOSE VIALS • CODE 4318-JQUANTITY CONTENTS CODE1 Alkalinity Unit Dose Vials, 20 pouches 4318-JEquipment needed but not supplied:STAN

Pagina 271 - SILICA – LOW RANGE

PROCEDUREUse COD/UDV adapter1. Press and hold until colorimeter turns on. 2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another seq

Pagina 272

ALUMINUMERIOCHROME CYANINE R METHODCODE 364I-01-SCQUANTITY CONTENTS CODE5 g *Aluminum Inhibitor Reagent *7865-C2 x 120 mL *Aluminum Buffer Reagent

Pagina 273

 STATISTICAL & TECHNICAL DEFINITIONS RELATED TO PRODUCT SPECIFICATIONSMethod Detection Limit (MDL): “The method detection limit (MDL) is defi ne

Pagina 274

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 002

Pagina 275 - SILICA – HIGH RANGE

AMMONIA NITROGEN - LOW RANGESALICYLATE METHOD • CODE 3659-01-SCQUANTITY CONTENTS CODE60 mL *Salicylate Ammonia #1 *3978-H10 g *Salicylate #2 *7457

Pagina 276

PROCEDURE - FRESH WATER1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence

Pagina 277 - SULFATE – HIGH RANGE

1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Select ALL TESTS (or another sequence containing 004 Ammonia-

Pagina 278

AMMONIA NITROGEN - HIGH RANGENESSLERIZATION METHOD •CODE 3642-SCQUANTITY CONTENTS CODE30 mL Ammonia Nitrogen Reagent #1 V-4797-G2 x 30 mL *Amm

Pagina 279 - SULFIDE – LOW RANGE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select TESTING MENU.3. Scroll to and select ALL TESTS (or another sequence

Pagina 280

AMMONIA NITROGEN, High Range SMART3 Test Procedures 11.10Test Procedures

Pagina 281

BARIUMBARIUM CHLORIDE METHOD • CODE 3638-SCQUANTITY CONTENTS CODE50 g Barium Reagent Powder 6330-H1 Spoon, 0.5g, plastic 0698*WARNING: Reagents marke

Pagina 282

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence containing 018

Pagina 283 - SURFACTANTS

BENZOTRIAZOLE/TOLYLTRIAZOLEUV Photolysis Method • CODE 4047QUANTITY CONTENTS CODE15 g *Benzotriazole Reagent *3818-E25 mL Potassium Sodium Tartr

Pagina 284

5 Jeffery G. H., Basset J., Mendham J., Denney R. C., Vogel’s Textbook of Quantitative Chemical Analysis, 5th ed., Longman Scientifi c & Technical,

Pagina 285 - CODE 3666-01-SC

APPLICATION: Corrosion and rust inhibitors in cooling water systemsRANGE: 0.0 – 30.0 ppm Benzotriazole0.0 – 30.0 ppm TolyltriazoleMDL: 0.5 ppm Benzotr

Pagina 286

BENZOTRIAZOLE PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence

Pagina 287 - TURBIDITY

TOLYLTRIAZOLE PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence

Pagina 288

BIGUANIDECOLORIMETRIC METHOD • CODE 4044QUANTITY CONTENTS CODE2 X 60 mL Biguanide Indicator 3994-H1 Pipet, plastic, 1.0 mL 0354Biguanide is a non-chlo

Pagina 289 - ZINC – LOW RANGE

PROCEDURE1. Press and hold until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence containing 006

Pagina 290

BORATE - UDVUNIT DOSE VIALS • CODE 4322-JQUANTITY CONTENTS CODE50 Borate UDV, 20 pouches 4322-JEquipment needed but not supplied:STANDARD ACCESSORY PA

Pagina 291

APPLICATION: swimming pools, surface and saline waters, hydroponic solutions, industrial waste. RANGE: 0.00 – 80.0 ppm boronMDL: 5 ppmMETHOD: Borate r

Pagina 292 - APPENDIX

PROCEDUREUse COD/UDV adapter1. Press until colorimeter turns on.2. Press ENTER to select Testing Menu.3. Select ALL TESTS (or another sequence conta

Pagina 293 - Appendix

BORATE, UDV SMART3 Test Procedures 11.10Test Procedures

Pagina 294

BORONAZOMETHINE-H METHOD · CODE 4868-01QUANTITY CONTENTS CODE120 mL *Boron Buffer *4869-J10 g *Boron Indicator Powder *4870-D1 Pipet, plasti

Commenti su questo manuale